Thermocouple Signal Converter

SIM µ 2.0 Thermocouple (TC) input type modules can be interfaced to any of the commonly available TC types; i.e. J,K,B,E,R,S,T,N to generate isolated 0-10V/4-20mA output. These modules comprise of in-built Cold Junction Compensation (CJC) calibrated to track the TC mV response at room temperature. The TC mV v/s V/I output is a linearly proportional value. The module can be calibrated to I/P v/s O/P range according to the user requirements.

The user has the option of implementing non-linearity error compensation of the TC response in the Data Acquisition System interfaced to the Voltage/Current output of the modules to achieve better accuracy in temperature measurements.

Ordering Information

Isolation (I) Input Type Output Type Supply I/P Signal Range O/P Signal Range Isolation
I I/P O/P Supply XX X 3P
I TC V
I
2D Type J :XJ ( 0 – +1200º C)*
0-5V:05 3P
Type K :XK ( 0 – +1370º C)*
Type E :XE
(0 – +1000º C)*
0-10V:10
Type B :XB
( 800 – +1820º C)*
Type R :XR
( 0 – +1750º C)*
4-20mA:42
Type S :XS
( 0 – +1750º C)*
Type T :XT
(0 – +400º C)*
0-20mA:20
Type N :XN
( 0 -+1300º C)*

*: Max. Measurement Range; X: Actual Range of TC

 

Download PDF

  • Input Thermocouple types : J, K, B,E,R,S,T,N
  • In-built Cold junction compensation (CJC).
  • 0-10V/4-20mA output calibration provided as per user requirement of input temp. range.
  • Thermocouple non-linearity not compensated
  • 24VDC Power Supply.
  • Slim & compact design.
  • Energy Efficient Module.
  • Low Noise & Low drift for long term stability
  • 2kV 3 Port Isolation.

Applications

  • Heating Control Systems
  • Temperature monitoring systems
  • Data Acquisition Systems

Download PDF

 

Thermocouple Input Signal Input Options: J Type(0 to +1200°C)* Factory Set
K Type(0 to +1370°C)*
E Type(0 to +1000°C)* * Max. Temperature Range
B Type(800 to +1820°C)*
R Type(0 to +1750°C)*
S Type(0 to +1750°C)*
T Type(0 to +400°C)*
N Type(0 to +1300°C)*
Input Impedance 100 k Ohm
Output Signal Output Options: 0- +5V, 0- +10V, 0-20mA, 4-20mA. Factory Set
Load Resistance For Output Voltage >1 K Ohm
For Output Current 500 Ohm Max.
Performance
a. Electronic System
Accuracy ±0.2% of Span at 25°C
Linearity 0.1%
Temperature Drift <0.01% / °C
Response Time < 1Sec
Calibration Zero & Span (Trimpot )
b. Thermocouple Non-linearity Compensation : Not provided

Download PDF

Isolation 2KV – 3 Port.
Power Supply Voltage 24VDC(nominal)
Operating Range 18 to 36 VDC
Ripple < 100mVpp
Power Supply Current 150mA.Max
Enclosure ABS Plastic, 35 mm DIN Rail Mounted. IP40 Protection class. Single Output W – 22.5mm
H – 75mm
D – 110mm
Operating temp. range 0 – 55°C
Storage temp. range 0 – 80°C
Relative Humidity 0 – 90%
RH Non-Condensing
Max. Altitude 3000 mts above MSL

 

Download PDF

Single input & Single output Converter

Pin Details Pin Details
1 +Input 5 +Output
2 -Input 6 -Output
3 NC 7 +24Vdc
4 NC 8 GND

 

Download PDF

Additional information

Input Types

Thermocouple Sensor

Output Signal

Current 0-20/4-20mA (4 Wire), DC Voltage (+/-10V)

Enclosure

ABS Plastic

Product Type

Converter

Thermocouple Type

Thermocouple Signal Converter

Reviews

There are no reviews yet.

Be the first to review “Thermocouple Signal Converter”

Your email address will not be published. Required fields are marked *